Immediately after each use, the wand should be cleaned by soaking in enzymatic detergent for 15 minutes. The wand should then be washed in hot running water and the abrasive surface scrubbed with a brush to remove the protein material.
The wand should then be placed in a disinfecting solution. Read the label on the solution for the proper soak time.
STEAM STERILIZATION:
The wand can withstand steam sterilization. Steam sterilization should always be used if the surface of the skin may be penetrated. If steam sterilized, disinfectant soak is not necessary.
The DiamondTome™ MicroExfoliation wands are reusable and should be cleaned after each treatment. The equipment should be steam sterilized** if the surface of the skin has been penetrated.
Gravity Displacement Steam Sterilization:
If the wand is sterilized in an unwrapped tray the “flash” cycle may be used. This is five minutes at a minimum temperature of 132°C (270°F) in a steam autoclave.